Baud Rate
Data bits
Parity
Stop bits
Quantities...
Display
Resolution
Service Info
Large Font
Harm. Graph...
Contrast...
VGA Output
Beep
Lock@SwitchOn
Lock Password
Lock Now ->
FoutX Config. ...
Fin Meas/Test ...
OPS Meas/Test ...
16
3.2.14 RS-232 Menu
- parameter setting the RS-232 serial line communication rate set value in bauds.
The value of the parameter is 4800, 9600, 19200, 38400, 57600 or 115200
bauds, respectively,
- parameter setting the number of data bits set value on the RS-232 serial line.
The parameter value is 7 or 8, respectively,
- parameter setting the RS-232 serial line set value. The parameter value is
"ODD" for the odd parity, EVEN for the even parity and NO for the no parity,
respectively,
- parameter setting the number of stop bits set value on the RS-232 serial line.
The parameter value is 1 or 2, respectively,
3.2.15 Display Menu
- quantities menu
- parameter setting the state of measured results values appeared on the display.
The value of the parameter is ON or OFF for the switched-on or switched-off
results depicting on the display, respectively. Parameter is not related to the
device menu depicting, which is displayed always if it is activated,
- resolution setting: STANDARD = results with standard number of decimal digits
according to the shown variable type; HIGH – number of decimal digits is raised
by 1 comparing to option STANDARD; MAXIMAL - number of decimal digits is
raised by 2 comparing to option STANDARD.
- enabling or disabling (ON / OFF) of service information
- parameter setting the state when measured quantity is displayed in large font.
The option is activated if only one quantity is chosen to be displayed,
- harmonic items graph menu
- display contrast change menu
- parameter Disables or Enables VGA output on the rear side of newer devices
3.2.16 Keyboard Menu
- enabling or disabling the beep functionality of the keyboard,
- enabling or disabling the automatic keypad lock right after each device switch
on
- keypad unlocking password change
- instant keypad lock with a basic mode activation.
3.2.17 Impulse Menu
- four similar items (where X = 0,1,2,3) for four impulse outputs of the device.
Each one represents an access into submenu containing settings of particular
impulse output.
- Frequency Input menu
- OPS Measurement Mode menu
RS-UG91-EN
Reference Standard RS 2x30, RS 1x30
User's Guide