11 Instant help

In case of an error in the program process, briefly turn off the appliance and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch device off and then on again. If the error
message remains inform manufacturer.
39
Possible cause
The display unit is not switched on.
Mains power supply interrupted (mains cable
defective).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
Draught/air movement
Table/floor vibrations
Weighing platform has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The display of the balance is not at zero
Adjustment is no longer correct.
The weighing pan is not level
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
KDE-T-BA_IA-e-1210