In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.
Failure:
The displayed weight
does not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
20
Possible causes:
 The display unit is not switched on.
 The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
 Power supply interrupted.
 Rechargeable battery inserted incorrectly or empty.
 No rechargeable battery inserted.
 Draught/air movement
 Table/floor vibrations
 The load cell has contact with foreign objects
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
 The display of the balance is not at zero.
 Adjustment is no longer correct.
 Great fluctuations in temperature.
 The weighing system is not on even surface.
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
KMF-TM-BA_IA-e-1310