Section 8-7 Power Block
8-7-1
Transformer Assembly (FRU No. 5196448)
This is a description on how to remove and replace the Transformer Assy.
8-7-1-1
Tools
Common Phillips screwdrivers
8-7-1-2
Needed Manpower
1 persons, 15 minutes + travel
8-7-1-3
Preparations
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-7-1-4
Removal Procedure
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
CAUTION
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
1.) Remove Rear cover
2.) Unscrew two screws (a,b).
3.) Remove GND cable (which ever connected near to "a" )
4.) Pull out the transformer tray so the connector marked J4 is accessible and disconnect Cable
(J4TRx)
8-64
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 213 of 252
3
EVISION
a
Figure 8-53 Removing Transformer Assembly