8-3-4
Optical Trackball Assy (FRU P/N # 5262338)
Tools
8-3-4-1
Common phillips screwdrivers
8-3-4-2
Needed Manpower
1 person, 10 minutes + travel
8-3-4-3
Preparations
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-4-4
Removal procedure
1. Remove Keyboard Assembly.
2. Unfasten the 3 screws holding the trackball bracket (circled in cyan color in Figure 8-12)
3. Remove Trackball assembly with Trackball Interface cable.
8-3-4-5
Functional Checkout Procedure
Functional and/or
Leakage Current
See Section
4-3-1
Power On/Boot Up
4-4-1
Basic Controls
4-4-2
B- Mode Checks
4-4-3
M-Mode Checks
4-4-4
Measurement Check
Section 4-5
Peripheral Checks
Section 7-6
Service Diagnostics
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 164 of 252
3
EVISION
Figure 8-12 Optical Trackball Assembly
Test
Service Manual Direction 5213326-100, Rev 3, Section 8-3-4. Equipment passes all
3 screws to be
removed
Debrief Script
required tests and is ready for use.
Chapter 8
LOGIQ™ A3 S
M
ERVICE
ANUAL
8-15