GE LOGIQ A3 Basic Service Manual page 180

Table of Contents

Advertisement

GE H
EALTHCARE
D
5213326-100, R
IRECTION
8-4-6
Rear Arm Cover Set (FRU P/N # 5263051)
Tools
8-4-6-1
Common phillips screwdrivers
8-4-6-2
Needed Manpower
1person, 15 minutes + travel
8-4-6-3
Preparations
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-6-4
Removal Procedure
1. Remove Keyboard rear cover.
2. Remove Rear Arm cover set. Retain the fasteners for the rear arm covers.
RH Arm
Cover
8-4-6-5
Functional Checkout Procedure
Functional and/or
Leakage Current
See Section
4-3-1
Power On/Boot Up
4-4-1
Basic Controls
4-4-2
B- Mode Checks
4-4-3
M-Mode Checks
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 180 of 252
3
EVISION
Figure 8-25 Arm Covers
Test
Service Manual Direction 5213326-100, Rev 3, Section 8-4-6. Equipment passes all
LH Arm Cover
Debrief Script
required tests and is ready for use.