DAKOTA ULTRASONICS MMX-7 Operation Manual page 61

Data logging thickness gauge
Hide thumbs Also See for MMX-7:
Table of Contents

Advertisement

1) Assuming the ALARM STATUS is ON,use the UP and DOWN arrow keys to
scroll through the sub menu items until ALARM HI LIMIT is highlighted.
2) Press the LEFT and RIGHT arrow keys to scroll the value. When the
correct alarm value is being displayed, proceed to step 8 .
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the ALARM HI value is correctly displayed.
7)
Press the OK key to set the ALARM HI value and return to the menu
screen, or ESC to cancel entering the ALARM HI value.
8)
Press the MEAS key to return to the measurement screen and begin taking
readings.
MMX-7 Data Logging Ultrasonic Thickness Gauge
Setting the Alarm HI Limit
57