Scale Adjustment; Using Markers - Protek A434L Operating Manual

4 ghz cable & antenna analyzer
Table of Contents

Advertisement

Action
Make a measurement after completion of O-S-L calibration.
1. Connect the cable to measure its
loss to the RF out port of the
instrument.
2. Connect the Short standard of the
Cal Kit to the end of the cable to be
tested
Table 4.6 Cable Loss Measurement Procedure

4.3.4.1 Scale Adjustment

Press the Auto Scale key to optimize the Y scale and display an
entire trace.
Press the AMP key to set the maximum and minimum values on
the Y scale manually.
Press the Scale key to select a display unit of the Y scale.

4.3.4.2 Using Markers

Depending on the draw type setting, four or six markers can be used
simultaneously. The level of a current trace is displayed next to each marker
on the screen, but the frequency information is not provided. Activate a
marker table to see both level and frequency information when moving a
marker to the specific frequency using the Marker Edit.
96
Note
Cable Loss measurement result is
displayed on the screen.