Cable Loss Measurement - Protek A434L Operating Manual

4 ghz cable & antenna analyzer
Table of Contents

Advertisement

6. Connect the Load
standard.
7. Select the
Load/Start
screen
menu.
Calibration state is changed to "Cal On" after the Open-Short-Load calibration.
4.3.4

Cable Loss Measurement

The instrument is ready to perform Cable Loss measurement after
completion of Open-Short-Load calibration.
Figure 4.14 One Port Cable Loss Measurement Connection Diagram
The instrument must be calibrated to get reliable Cable Loss measurement
results. For best results, set the frequency and calibrate the instrument
immediately before taking measurements.
Connect CAL Kit "Load" connector to the RF out port.
When the
progress bar is displayed to show the progress.
When a progress bar is disappeared, Load
calibration is finished.
Table 4.5 Calibration Procedure
Load/Start
screen menu is selected, a
95