Clarus SQ 8 MS Hardware Guide
Problem
Problem
Poor or inadequate
sensitivity.
134
Probable
Probable Cause
Cause
Leaking injector septum.
Replace the septum.
Leak from injector
Tighten/replace ferrules.
ferrules.
Foreign material in the
Clean the injector.
injector.
Peak splitting.
Prevent double injections.
Dry the outside of the injector
needle.
Replace the injector septum.
Injector and column are
Install a silanized injector
more active toward
liner, or silanize the current
acid/base compounds.
injector liner.
Check or replace the injector
packing material, such as
quartz wool.
Cut off the first 30 cm of the
column and rerun the test mix.
If the results do not improve
replace the column.