Clarus SQ 8 MS Hardware Guide
Problem
Problem
Turbo pump will not
accelerate
Vacuum light continues to
blink.
High mass spectra appears
as a large blotch, or loss of
high mass spectra.
Drastic change in mass peak
shape for no apparent
reason.
No spectra, or large blotch.
132
Probable
Probable Cause
Cause
Pump malfunction.
Call a PerkinElmer service
engineer.
Pump controller
Call a PerkinElmer service
malfunction.
engineer.
GC is not properly
Set the proper GC
configured.
configuration for your site.
Large leak.
Locate vacuum leak and
correct.
Foreline trap has
Replace filter.
excessive moisture.
Rotary pump set to gas
Switch the gas ballasting off.
ballast.
Rotary pump requires oil
Change oil.
change.
Bad tune.
Run AutoTune.
Bad tune.
Run AutoTune.
Bad tune.
Run AutoTune.
Solution