EM TEST compact NX5 Series Operation Manual page 98

Ultra-compact simulator and its system modules
Table of Contents

Advertisement

EM TEST
10.3.1.2. Start Trigger
The Start Trigger occurs by pressing the blinking START / PAUSE button or with an external trigger
positive slope signal at the rear side BNC Ext. Trigger plug.
The start trigger offers the following status:
- Automatic
- At Start (starts after the trigger)
- Each pulse (same as trigger out)
- Each step (same as trigger out)
If the start trigger is not set to Automatic, the trigger indicates with "Waiting for trigger" the recommended
trigger signal.
The waiting time for a trigger signal is endless if no trigger timeout is enabled.
External trigger timeout
This function stops the test after the defined timeout time
and indicates the message, "Error trigger unit timeout".
For enable the function External trigger timeout select from
the Home screen:
Menu / Setup / General / Process & sync
Operating Manual
Start Trigger flowchart example from burst
V 1.06
Compact NX5
98 / 143