EM TEST
3.2.5.
Mains Switch and fuse
The mains power voltage indicated on the instrument must correspond with the local supply voltage (mains
voltage: 85–265 Vac, universal power unit, mains frequency: 50–60 Hz).
Mains Switch, fuse holder and power input
3.2.6.
Connecting the compact NX5 system to the ground reference
Connection to reference ground
Earth bolt dimension
Operating Manual
To replace a fuse:
1) Disconnect the mains cable
2) Pull the fuse holder out of the connector
3) Remove the damaged fuse(s)
4) Insert 1 or 2 fuses (4 A / 115V and 2 A / 230 V slow blow)
5) Replace the fuse holder
6) Plug the mains cable into a power outlet with a solid ground
connection
7) Switch the system on and operate as instructed in this manual
Connection to reference ground
For burst tests, the generator must be placed on a ground
reference plane which is connected to ground.
A low inductive high frequency ground connection between the
test system and the ground reference plane (GRP) is absolutely
essential for performing burst tests correctly.
Earth bolt dimensions:
Screws:
Metric M4 ,
Distance: 10.1 mm x 11 mm
Banana plug, 4 mm
Plug:
V 1.06
Compact NX5