Pulsed Magnetic Field As Per Iec 61000-4-9 - EM TEST compact NX5 Series Operation Manual

Ultra-compact simulator and its system modules
Table of Contents

Advertisement

EM TEST
6.3.

Pulsed magnetic Field as per IEC 61000-4-9

Test Level
Generator charging voltage
Short circuit current
Wave shape
Rise time tr
Pulse duration
Polarity
Repetition rate
Events preselection
Counter
Trigger
Trigger of pulses
Synchronization
Resolution
Measurements
CRO
CRO V
CRO I
Accuracy CRO V CRO I
CRO V/I Z measuring instrument > 1 MΩ
Peak voltmeter
Peak current meter
Current limiter setting
Output
Direct HV-COM
Coupling network
DUT supply
Test Routines
Quick Start
Standard test routines
separated to (AC / DC)
Extended Tests
Magnetic field test
* depends on charging voltage
Operating Manual
Model NX5
160 V – 5000 V ± 10% Step 1 V
80 A – 2500 A ± 10%
8 s ± 20%
20 s ± 20%
Pos., Neg., Alt
max. 1 Hz (1 s* - 999 s)
1 - 99'999
1 – 1'000'000
AUTO, MAN, EXTERN
0 - 360° (16 – 500 Hz)
Asynchronous = 0°
5 V Trigger
10 Vp at 5 kV
10 Vp at 2.5 kA
± 10%
5000 V
2500 A: Display  1 Digit  20 A
10 A...3000 A step 10 A
HV-Banana connector, Zi = 2
L – N with Z = 2
L-PE, N-PE, L+N-PE with Z = 12
AC 300 V / 16 A / 50/60 Hz
DC 300 V / 16 A
Immediate start, all parameters adjustable during a running test
IEC 61000-4-5 one level: Selectable level 1 to 4
IEC 61000-4-5 all levels: Starts from lowest level to selected level
IEC 61000-4-5 level 3
IEC 61000-4-5 level 4
EN 61000-6-1 Generic (AC / DC)
EN 61000-6-2 Generic
Voltage iteration stepwise: Voltage level V after n pulses by V
Angle iteration stepwise: Change phase angle A after n pulses by A
Angle random iteration: Random phase angle A after n pulses
test routine as per IEC 61000-4-9
test level 100, 300 and 1000 A/m
cont. adjustable within Quick Start
V 1.06
Approx. 80 A/m to 2500 A/m
a reference signal is connected to the Sync input.
no
x
Compact NX5
9 F
18 F
x
x
x
71 / 143