Power Frequency Magnetic Field As Per Iec 61000-4-8 - EM TEST compact NX5 Series Operation Manual

Ultra-compact simulator and its system modules
Table of Contents

Advertisement

EM TEST
13.

Power frequency Magnetic Field as per IEC 61000-4-8

Main Menu Magnetic Field separated to
"Power frequency" and "Pulsed"
Required Device settings to perform a power frequency magnetic field test
For perform a 50/60 Hz magnetic field test, the NX5 generator must be configured with the used hardware in the
menu:
Menu / SETUP / Equipment
1. Variable transformer:
- variac NX1-260-16
- variac NX1-260-32
2. Magnetic current transformer
- MC 26100
- MC 2630 (1 A range)
- MC 2630 (30 A range) H-Field
Note: The MC 2630 has two ranges and therefore the NX generator list this current-transformer in the
configuration for (1 A) and (30 A) range. Set the range switch of the MC 2630 to the correct position before
start the test.
Menu / SETUP / EXTERNAL COUPLERS
3. Magnetic field antenna:
- MS 100N
Set the used equipment to default status
Operating Manual
H-field 75 to 1100 A/m
H-field
1 to 9 A/m
9 to 75 A/m
V 1.06
Compact NX5