Magnetic Field Test As Per Iec 61000-4-9 - EM TEST compact NX5 Series Operation Manual

Ultra-compact simulator and its system modules
Table of Contents

Advertisement

EM TEST
11.

Magnetic field test as per IEC 61000-4-9

Pulsed Magnetic field as per IEC 61000-4Surge Module 1.2/50s – 8/20s
Main Menu Magnetic Field
The internal coupling network is designed for mains frequency 50 Hz / 60 Hz.
When L – N coupling is selected an additional current of approx. 1.5 A flows, caused by the
18 F coupling capacitor and the pulse forming network. A similar current appears during the
couplings to PE.
Tests with 400 Hz mains frequency destroy the coupling network. An external coupling
WARNING
network is recommended with inductors and capacitors matched to the 400 Hz mains
frequency.
Operating Manual
V 1.06
Compact NX5
105 / 143