VERIFICATION PROCEDURE
Verify that the remote can control the module
wirelessly (by increasing and decreasing the
intensities).
Verify that wireless communication is effective
in any location reachable with the arm and the
remote in the hand.
Verify that continuous transmission is operating
up to a distance of 2 meters.
Note: Final accessories may differ from the picture
By taking a pod off the body load during the
stimulation session, check that the stimulation
stops in case of electrode fault.
Note: Final accessories may differ from the picture