Intelect
HPL7
®
and HPL15
User Manual
1
EN

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga Intelect HPL7

  • Page 1 Intelect HPL7 ® and HPL15 User Manual...
  • Page 2 Images Handpiece 7” Touch screen display Emergency stop button Multi-function knob Calibration socket Label showing laser data Product identification label Main switch 2x2AT Fuses 100-240 V ~ 50/60 Hz socket Cooling fan Interlock Footswitch socket Interlock for patient button (Optional)
  • Page 3 Accessories Laser handpiece Ø 38 mm Applicator Ø 26 mm Applicator Magnetic attachment Magnetic attachment (207902) (207901) Footswitch Safety glasses Power cable (4930) (4800) (4910) 3 pole interlock (4900)
  • Page 4 Symbology on label and device Closed (connection of mains Sticker for footswitch socket supply) Open (disconnection of mains Label for interlock supply) Patient button interlock Applicator for optical fibre (only ScanX version) Product identification label Laser hazard (Intelect HPL 7) Attention: Laser Device of Class I, BF type specifications of IV class...
  • Page 5: Table Of Contents

    Contents 1. Introduction ..............7 1.1 What is Intelect HPL 7 and HPL15? .........7 ® 1.2 What is Intelect HPL 7 and HPL15 used for? ....7 1.3 Advantages ................7 1.4 Who can use Intelect HPL 7 and HPL15? ..... 8 1.5 Indications ................. 8 1.6 Contraindications ..............
  • Page 6 4. Warnings ..............25 4.1 Attention during use ............25 4.2 Therapeutic care ..............26 5. Standards and safety ..........28 5.1 Laser standards ..............28 5.2 Safety features ..............28 5.3 Classification .................30 6. Maintenance ...............32 6.1 Ordinary maintenance ............32 6.2 Troubleshooting and periodical maintenance ..33 6.3 Alarms ..................34 7.
  • Page 7: Introduction

    1. Introduction 1.1 What is Intelect ® Intelect HPL 7 (REF 2879) and HPL15 (REF 2979) are medical devices for high energy laser therapy, designed for the treatment of pathologies HPL 7 and HPL15? affecting muscles and bones such as lumbago, epicondylitis, shoulder pain, arthrosis, as well as contracture, sprains and many other complaints.
  • Page 8: Who Can Use Intelect Hpl 7 And Hpl15

    1.4 Who can use Intelect HPL 7 and HPL15 are designed to be used for laser therapy. Intelect HPL 7 and Laser therapy falls within the field of physiotherapy and analgesic therapy. HPL15? The operator must be qualified and competent and in one of the following categories: Medical personnel: Physiatrists...
  • Page 9: Contraindications

    1.6 Contraindications Because the laser source used emits light in the near infrared range, contraindications are essentially of two types. - Eye damage - Thermal damage to tissue exposed to the therapy, with particular attention to continuous emission and the phototype of the patient. The first case is prevented by using protective glasses, while the second is avoided through the operator’s ability to evaluate the correct thermal sensitivity of the patient and their phototype.
  • Page 10 Intelect HPL 7 and HPL15 isn’t suitable for use in areas of explosive, flammable or combustive atmosphere. The unit must be disconnected before every maintenance or clean, removing the cable from the socket. This is to avoid the risk of electric shocks.
  • Page 11: Quick Use Guide

    2. Quick use guide Before using the Intelect HPL 7 and HPL15, make sure it is resting on a stable surface. Make sure the power switch is set to “0”. Make sure the power cable is attached to the machine; connect the plug to a power socket. In case of any questions during the setup of the Intelect HPL 7 and HPL15, please contact DJO (contact list at the end of this User Manual).
  • Page 12 After the welcome screen, a keyboard will be displayed. Enter the password to start up the machine. The password is 1-2-3-4. Select the mode to start the therapy. Attach the magnetic applicator to the handpiece. Make sure that those present (operator and patient) wear certified protective glasses.
  • Page 13 Identify the area of the body to be treated and position the handpiece perpendicular to the area. To set the treatment parameters: press “Quick start”. This will bring you to a screen that allows you to choose the mode (continuous, interrupted, pulsed, single pulse, burst, stochastic, anti- inflam) before the emission starts.
  • Page 14: How To Use The Handpiece

    To reset the parameters (time, amount of energy and power), press the “stop” button. Turn off the machine using the switch on the back of the machine. 2.1 How to use Always use the laser handpiece with the applicator supplied. the handpiece We recommend performing the calibration procedure prior to every treatment.
  • Page 15: Instructions For Use

    3. Instructions for use After the device is switched on and the welcome screen has appeared, you will come to the main ‘device management’ screen, and will need to enter the password. 3.1 Start-up To start the treatment press “Quick start”, choose the emission mode (continuous, interrupted, etc), set power parameter by pressing “Laser power”...
  • Page 16: Mode

    3.2 Mode The modes are divided in 7 sections: A) Continuous (cap 3.2.1) B) Interupted (cap 3.2.2) C) Pulsed (cap 3.2.3) D) Single Pulse (cap 3.2.4) E) Burst (cap 3.2.5) F) Stochastic (cap 3.2.6) G) Anti-Inflam (cap 3.2.7) 3.2.1 Continuous mode In this part of the menu it is possible to select continuous mode functioning.
  • Page 17: Interrupted Mode

    3.2.2 Interrupted mode To start the treatment choose Interrupted emission mode, set power parameter by pressing “laser power” and energy quantity by pressing “Joule”. The duty cycle can be modified. Take out the handpiece from its holder, put on the handpiece spacer and press the footswitch to activate laser emission. The lower part of the screen shows the actual power delivered, the time, the amount of energy transferred and the temperature of the laser head.
  • Page 18: Single Pulse Mode

    3.2.4 Single Pulse mode This mode lets you set the duration of a single pulse. The Single Pulse mode delivers a pulse, the duration of which is defined by the operator each time the footswitch is pressed. It is possible to select a duration of 20 ms to a maximum of 1000 ms by pressing the square at the center of the screen.
  • Page 19: Stochastic Mode

    3.2.6 Stochastic mode Stochastic emission mode generates variable power laser pulses; this particular mode interacts with the peripheral nervous system, greatly limiting the perception of heat during the treatment. “Naturally thermal” emissions combined with thermo-mechanical action on the peripheral receptors, allow a rapid depolarization of nociceptive pathways, polarizing the A-delta fibers, and so counterposing the pain coming from noxa through the action of the Gate Control.
  • Page 20: Cps

    3.3 CPS The pathology section allows the operator to follow the guided treatment protocols. In this section there are a number of therapeutic protocols divided according to the anatomical area that you want to treat. It is possible to select the anatomical area from the displayed human body figure; pressing on the chosen area accesses the respective treatment protocols.
  • Page 21 The protocol, with automatic settings, will display values of power, emission mode, time and / or dosimetry for cm2 for each preset phase. By removing the handpiece from its holder, putting on the applicator, pressing “Start” and acting on the pneumatic footswitch, you can begin the therapy session. The lower part of the screen shows the actual power delivered, the time, the amount of energy transferred and the temperature of the laser head.
  • Page 22: Favorites

    3.4 Favorites In the “Favorites” section, you can enter and start protocols previously saved in the “Programming” screen. It is possible to start treatment after selecting a protocol. 3.5 Guide This section displays the therapeutic indications, and contraindications, and some information on the proper use of the device. 3.6 Programming In this section you can set favorite protocols, first selecting the output mode, and then the treatment parameters (time, power, joules etc) following the...
  • Page 23: Calibration

    Once you have chosen the parameters click “save” in the bottom right corner of the screen. Enter the name of the protocol or the patient then select where to save the protocol and click “save”. The treatment protocol will be saved in Favourites. 3.7 Calibration IEC EN 60601-2-22 (norm governing the safety of therapeutic and diagnostic laser equipment) states that: ‘The laser equipment must be...
  • Page 24: Settings

    3.8 Settings Through this screen it is possible to access product settings. Language: it is possible to set the software language. Buzzer Volume: it is possible to adjust the audible feedback emitted during the use of the product (treatment). Beep: In ‘Beep’ it is possible to switch touchscreen button sounds on or off. Maintenance: Add a password for the internal management of the device (reserved section).
  • Page 25: Warnings

    4. Warnings 4.1 Attention During use of Intelect HPL 7 and HPL15 it is necessary to take some during use precautions to keep the machine in good working order. In particular, it is necessary to take care of the tube which connects the handpiece to the machine: any damage could stop the machine working correctly.
  • Page 26: Therapeutic Care

    Pay attention to the trolley wheels: avoid running over the cable. Never turn off the Intelect HPL 7 and HPL15 using the emergency stop button! It is designed for emergency use only; its repeated use to switch off the laser may damage the machine. If the emergency stop button is pressed, return it to its normal position by turning it, to allow the...
  • Page 27 Do not mark the areas to be treated with ink, mark the outline of the area within which the applicator will be positioned. Ensure that the surface of the skin is clean and, most importantly, is not wet or sweaty. Remember that thermal sensitivity varies from patient to patient, and from area to area on the same person, and that the level of pigmentation greatly influences the absorption of energy and therefore the emission times at...
  • Page 28: Standards And Safety

    5. Standards and safety 5.1 Laser standards Intelect HPL 7 and HPL 15 have been designed and built in compliance with the following standards: EN 60601-1 IEC 60601-2-22 EN 60601-1-2 EN 60601-1-6 EN 62366 EN 62304 IEC 60825-1 5.2 Safety features Intelect HPL 7 and HPL15 is a class IV laser and as such features a safety interlock.
  • Page 29 Laser warning It is recommended that the machine is used in a well-lit room, without reflective surfaces, with the laser warning sign displayed at the entrance. Laser warning light Protection There is a risk of heat damage to tissue exposed to radiation which cannot be removed without making treatment impossible, given the power involved and the focusing systems used in DJO devices.
  • Page 30: Classification

    5.3 Classification The lasers in the Intelect HPL 7 and HPL15 family are therapeutic and surgical lasers with optical fibre output, designed to function with pumped semiconductor laser heads, with wavelengths between 600nm and 2000nm, with maximum power up to 7W (Intelect HPL 7), up to 15W (Intelect HPL 15).
  • Page 31 The circuitry is generally oversized and well ventilated, which avoids overheating and associated problems. The recommended operating temperature is between 10 and 30 °C, with humidity between 30 and 75% and pressure between 700 and 1060 hPa. When used in a clinic or hospital, in a normal non-sterile environment, Intelect HPL 7 and HPL15 has no particular protection against accidental spillage of liquids;...
  • Page 32: Maintenance

    6. Maintenance 6.1 Ordinary The Intelect HPL 7 and HPL15 laser system does not require particular maintenance procedures; however it is recommended for use in areas which maintenance are free of dust and the machine should be cleaned carefully. To clean the screen and the case use a professional microfiber wipes for screen cleaning.
  • Page 33: Troubleshooting And Periodical Maintenance

    6.2 Troubleshooting Refer to the table below for troubleshooting. For problems of any other type and periodical please contact the DJO Assistance Centre. maintenance Problem Probable Cause Solution A) No power supply A) Check power supply Intelect HPL 7 / HPL15 does B) Blown fuse B) Change the fuse not switch on.
  • Page 34: Alarms

    6.3 Alarms In the case of malfunction the following warnings may be shown on the Intelect HPL 7 and HPL15 display. Message Probable cause Solution The laser electronics should be Contact the DJO CHECK MACHINE HOURS checked Assistance Centre Cool the laser keeping the machine switched on.
  • Page 35: Labels

    7. Labels Applicator for optical fibre Laser hazard Laser data plate Product identification plate (Intelect HPL 7) Sticker showing footswitch connection Sticker showing interlock connection for patient button (Optional) Interlock connection...
  • Page 36: Warranty

    8. Warranty DJO France ("Company") warrants that the Intelect HPL 7 and HPL15 system is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If these Products fail to function during the two year warranty period due to a defect in material or workmanship, at the Company's option, Company or the selling dealer will repair or replace the respective Product without charge within a period of thirty (30) days from the date on which...
  • Page 37: Technical Data

    9. Technical data Technology Real-Time Operating System Wavelenght 810+980 nm (HPL 7), 810+980+1064nm (HPL 15) Guide Light 650 nm - 3 mW Laser power 7 W (HPL 7), 15 W (HPL 15) Emission Continuous (CW), Pulsed, Single Pulse, Burst, and Stochastic (patented stochastic mode) Operation mode Continuous, Interrupted, Pulsed, Single Pulse, Burst, Stochastic, Anti-Inflammatory...
  • Page 38: Emc Tables

    10. EMC tables...
  • Page 42 DJO GLOBAL AUSTRALIA: BENELUX: CANADA: T: +1300 66 77 30 T: Belgium 0800 18 246 T: +1 1866 866 5031 F: +1300 66 77 40 T: Netherlands 0800 0229442 F: +1 1866 866 5032 E: customerservice.au@DJOglobal.com T: Luxemburg 8002 27 42 E: canada.orders@DJOglobal.com E: benelux.orders@DJOglobal.com CHINA:...