Quick Start Guide
2
Configuration
2.1
Configure the transmitter
You must configure the transmitter for certain basic variables to operate.
Manufacturers pre-configure the transmitters in the factory to order
specifications or factory defaults.
You may need to configure if the transmitter has not been configured or if
the configuration variables need revision. To configure, do one of the
following:
•
Order factory-configuration from Emerson Automation Solutions.
•
Use the Rosemount 148 PC Programming interface in a bench
configuration setting.
The Rosemount
and a communication modem.
Prerequisites
The device will need an external power supply of 12–42.4 Vdc.
Procedure
1. Hook up the transmitter and a load resistor (250–1100 ohms) wired
in series with the power supply.
2. Attach the modem in parallel with the load resistor and connect it to
the PC.
2.2
Verify configuration
If the transmitter has a sensor connected (either a test sensor or actual
installation hardware), you can check the configuration with the
Information tab on the Rosemount
Procedure
1. Select Refresh to update the status and confirm that the transmitter
has been configured correctly.
2. If there are any problems, refer to the Rosemount 148 Temperature
Transmitter
6
™
148 PC Programming Kit includes configuration software
Reference Manual
™
148 PC Programmer Interface.
for troubleshooting suggestions.
April 2019
Rosemount 148