13.2 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS
C 051 Thorax is intended for use in the electromagnetic environment specified below. The customer or
®
user of the ATMOS
Immunity Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surges
IEC 61000-4-5
Power Frequency 50/60
Hz
Magnetic field
IEC 61000-4-8
42
C 051 Thorax should ensure that it is used in such an environment.
®
IEC 60601-Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
± 1 kV Differential
± 2 kV Common
3 A/m
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV for mains cable
inapplicable
± 1 kV input and output
lines
± 1 kV symmetric
± 2 kV symmetric
applicable
3 A/m
Electromagnetic Envi-
ronment - Guidance
Floors should be wood,
concrete, or ceramcis tile.
If floors are synthetic,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency magne-
tic fields should be that of
a typical commercial
or hospital environment.