9.0 Maintenance and service

9.1

Basic information

The device is equipped with a sensitive pressure sensor. Thus, the vacuum unit should undergo a safety check
and should be calibrated respectively by the manufacturer or an authorised service centre every two years. This
check should also include a check of the battery capacity as well as all parts for wear or damage.
Further regular maintenance work is not required.
It is understood that the vacuum unit and its application parts are regularly and thoroughly cleaned and disin-
fected, and that it is operated in accordance with the instruction manual.
Please observe any national and international regulations applicable for your institution
9.2

Repairs

The following problems could require repairs by the manufacturer or an authorised service partner. Please
contact the manufacturer or authorised service partner by telephone before sending in the unit.
Liquid has penetrated the device;
Significant decrease of battery capacity;
Sudden occurrence of abnormal displays on the screen;
Sudden occurrence of abnormal noises;
Operational faults or malfunctions which cannot be remedied using the measures as stated under "Remedy
of operational faults and malfunctions".
Measures when sending in the vacuum unit:
If the vacuum unit must be sent to the manufacturer or an authorised service partner – after respective consul-
tation with such manufacturer or authorised service partner – please ensure the following:
Please send in the device complete with charger and mains cable (see scope of delivery).
Remove all single-use and/or consumables;
Thorough cleaning and disinfection;
Air-tight packaging;
Accompanied by a detailed error description.