Knob And Key Functions - VWR Micro Star 30R Instruction Manual

Laboratory centrifuge, inclusive angle rotor ra 24/2 and aerosol tight rotor lid/without rotor
Table of Contents

Advertisement

INSTRUCTION MANUAL
10
TIME key
11
Indicator for selection of braking level ( )
12
Program display
13
PROGRAM key
14
Knob
15
FAST COOL key
16
LID key
17
START key
18
STOP key

3.2 KNOB AND KEY FUNCTIONS

By rotating the knob, you change the values of the parameters on display. By
rotating the knob clockwise, the value increases, and by rotating it counter-
clockwise, the value decreases. By pressing on the knob, you confirm the new
value for the parameter.
(Knob)
By pressing this key, you can set the temperature. The key lights up. Set new
temperature by rotating the knob and confirm it by pressing on the knob. The
key light will then turn off.
(TEMPERATURE)
By pressing this key, you can set the rotational speed (RPM) or centrifugal
force (RCF). The key lights up. By repeated pressings on this key, you select
which parameter will be set. When indicator RPM is lit, you can set the
(SPEED)
rotational speed, and when indicator RCF is lit, you can set the centrifugal
force. Set new values of both parameters by rotating the knob and confirm
them by pressing on the knob. The key light will then turn off.
By pressing this key, you can set the time (TIME), acceleration level ( ) and
braking level ( ). The key lights up. By repeated pressings on this key, you
select which parameter will be set. When indicator TIME is lit, you can set the
(TIME)
run time, when indicator
indicator
parameters by rotating the knob and confirm them by pressing on the knob.
The key light will then turn off.
By pressing this key, you can set the programs. The key lights up. Select new
program by rotating the knob and confirm it by pressing on the knob. The key
light will then turn off.
(PROGRAM)
If you want to change the values of parameters of an existing program, set the
values of individual parameters as described above, when this key is lit.
By pressing this key, you turn on the fast cool function. The key lights up.
Rotor starts to run automatically with 8,000 RPM. Fast cool function cools
down the centrifugal chamber to the set temperature. When the temperature is
(FAST COOL)
reached, display for rotational speed shows message READY, and the key
light will then turn off.
is lit, you can set the acceleration level, and when
is lit, you can set the braking level. Set new values of all
EN 8 / 27