Temperature Correction; Maintenance And Care; Regular Service; Cleaning The Centrifuge - VWR Micro Star 30R Instruction Manual

Laboratory centrifuge, inclusive angle rotor ra 24/2 and aerosol tight rotor lid/without rotor
Table of Contents

Advertisement

INSTRUCTION MANUAL

3.19 TEMPERATURE CORRECTION

If you notice a difference between preset temperature on display and actual measured temperature,
the centrifuge has temperature correction functionality. Temperature correction is performed by
entering the temperature difference in the service parameters of the centrifuge.
Temperature correction may only be performed by an authorized service technician!
Measured temperature depends on a number of factors, like environment temperature, rotational
speed, run time, etc.

4. MAINTENANCE AND CARE

4.1 REGULAR SERVICE

We recommend having the centrifuge and associated rotors checked by authorized service at least
once a year. You must thoroughly clean and disinfect the centrifuge prior to service.

4.2 CLEANING THE CENTRIFUGE

After every centrifugation, please remove any condensed water from the rotor chamber using a
soft, absorbent cloth. After use, leave the centrifuge lid open, so the inside can dry out.
For regular cleaning of the outside surface of the centrifuge and the rotor chamber, use mild
neutral detergent. Make sure that no liquid penetrates the inside of the housing.
Open the lid of the centrifuge and turn off the main switch. Disconnect the power cord from
mains socket. Remove the rotor with rotor key. Clean all accessible surfaces of the device and
accessories at least once a week and every time, when contaminated.
After cleaning with detergent, the rubber seal around the rotor chamber should be thoroughly
cleaned with water and lubricated with glycerine, to prevent it from becoming brittle.
For cleaning and disinfection, use only neutral cleaners and disinfectants. Before cleaning or
decontaminating the centrifuge, using means and methods not recommended in this manual,
you should consult with the manufacturer, in order to avoid the damage to the centrifuge.
To ensure safe and long operation of the centrifuge, please avoid the use of aggressive
chemicals, which can damage the centrifuge, rotor and accessories. Please check them
regularly for damage caused by corrosion.
Cleaning intervals:
For the sake of personal, environmental, and material protection, it is your duty to clean and if
necessary disinfect the centrifuge on a regular basis.
Maintenance
Clean rotor chamber
Clean rotor
Accessories
Cabinet
Ventilation holes
Recommended interval
Daily or when contaminated
Daily or when contaminated
Daily or when contaminated
Once per month
Every six months
EN 17 / 27
VWR Micro Star 30R