Pre-Selecting Run Time, Acceleration And Braking Level - VWR Micro Star 30R Instruction Manual

Laboratory centrifuge, inclusive angle rotor ra 24/2 and aerosol tight rotor lid/without rotor
Table of Contents

Advertisement

INSTRUCTION MANUAL

3.9 PRE-SELECTING RUN TIME, ACCELERATION AND BRAKING LEVEL

By pressing the TIME key, you can set the time (TIME), acceleration level ( ) and braking
level ( ). The key lights up. By repeated pressings on this key, you select which parameter
will be set. When indicator TIME is lit, you can set the run time, when indicator
set the acceleration level, and when indicator
values of all parameters by rotating the knob and confirm them by pressing on the knob. The
key light will then turn off.
• Run time
Run time can be set between 0.10 and 99.5 minutes. Time setting from
0.10 to 9.59 minutes is possible in 1 second steps. Between 10.0 and 99.5
minutes the setting is possible in 10 second steps. By rotating the knob
either clockwise until after 99.5 is displayed, or counter-clockwise until after
0.10 is displayed, you can set continuous operation (HLd on display).
NOTE: 0.10 = 10 s, 9.59 = 9 min 59 s, 10.0 = 10 min, 99.5 = 99 min 50 s
• Acceleration level (AcX)
Acceleration level can be set from 0 to 9. Level 0 means very slow
acceleration, level 9 very fast acceleration.
• Braking level (brX)
Braking level can be set from 0 to 9. Level 0 means stopping without
braking, level 9 maximum braking.
NOTE:
All the operation parameters can be changed during the operation of the centrifuge, and it starts
to work with the new settings after that.
Changing of run time during the operation:
You should be aware, that when you extend run time, the difference between newly set and
originally set time is added to current time, and when you shorten run time, the difference is
subtracted.
Example: The centrifuge started with time set to 10 minutes. It has been running for 3 minutes.
Then you changed the time to 5 minutes. The centrifuge will run for another 2
minutes.
is lit, you can set the braking level. Set new
EN 13 / 27
VWR Micro Star 30R
is lit, you can

Hide quick links: