The SAS SUPER ISO 100 and SAS SUPER ISO 180 can be fixed to a table/floor tripod or a wall support. The screw
thread connection is located underneath the unit, between the two front feet.
Functions
Software function
Touch the screen for at least two seconds to switch the unit on.
The air sampler is supplied with eight programmable air volumes ("Select volume") stored in the memory and eight
sampler configuration ("Configuration") selectable by the operator.
The selection of the volume is made by pressing the "UP" or "DOWN" arrows when the programme in the relevant
sub menu.
The following volumes of air are suggested:
- Contaminated areas (communities, processing rooms, etc) 10 - 200 litres of air
- Normal areas (laboratory benches, houses, etc) 200 - 500 litres of air
- Sterile or high risk areas (cleanrooms, operating theatres, etc) 500 - 1000 litres of air

To start with the same air volume as the previous sample

Switch on the unit by touching the screen for at least two seconds. After the visual display presentation you will see
the following display with the last aspirated volume of air:
Date
Last aspirated 
volume of air
Press START on the screen to start air sampling with the last cycle. The motor will run until the figures in the
brackets reach the same figure as displayed to the right and then it will stop.
Time
08/31/06
11.31
500 Lt. 
DELAY 00 MIN
START
Battery status
Last delay 
time set
MENU
11