Table 1-1. Technical data VWR MICRO STAR 17
Feature
Noise level at maximum speed
Temperature range
Dimensions
Height
Width
Depth
Weight with empty rotor

Mains supply

The following table contains an overview of the electrical connection data for the VWR MICRO
STAR 17 / 17R. This data is to be taken into consideration when selecting the mains connection
socket.
Table 1-2. Electrical connection data
Cat.
521-1646
521-1647

Functions and features

The following table gives an overview of the important functional and performance characteristics of
the VWR MICRO STAR 17 / 17R.
Table 1-3. Functions and Features
Component / function
Structure / housing
Rotor chamber
Drive
Keys and display
Controls
Internal memory
Functions
Lid lock
MICRO STAR 17
< 56dB (A)
230 mm
240 mm
350 mm
10.5 kg
Mains
Frequency
voltage
230 V
50 / 60 Hz
230 V
50 / 60 Hz
Description / features
Sheet metal with attached plastic housing and steel chamber
Plastic
Up tp 48 ml of spilled liquid is retained in the chamber and cannot enter the instrument.
Induction drive without carbon brushes
Easy-to-clean keypad and display surface
Microprocessor-controlled
The most recent data is saved
RCF, temperature, and pretemp selection
Automatic lid closing and locking starting from an initial hold position
Power
Rated current
consumption
1.4 A
180 W
1.9 A
300 W
1
Introduction and description
MICRO STAR 17R
< 50 dB (A)
Adjustable from -9 °C to 40 °C, in 1°C
increments.
330 mm
292 mm
440 mm
28.0 kg
Equipment fuse
Building fuse
4A, 2-pole circuit
16 AT
breaker
4A, 2-pole circuit
16 AT
breaker
VWR MICRO STAR 17 / 17R
Mains supply
1-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Micro star 17r521-1646521-1647

Table of Contents