Contact plate and the Petri dish holders can be adjusted (using a hex key) in case the
available plates are slightly different in diameter from the standard 55 mm contact plate
or 90 mm Petri dish.
Petri head adapter (optional)
An adapter is available in aluminium or stainless steel which allows the use of standard
90 mm Petri dishes with SAS for contact plates.
Filling 90 mm Petri dishes
We recommend filling standard 90 mm disposable Petri dishes with no more than 18 - 20
ml of agar to avoid the medium touching the inside surface of the aspirating head.
Tripod installation (optional)
The SAS SUPER ISO USB 100 and SAS SUPER ISO USB 180 can be fixed to a table/floor
tripod or a wall support. The screw thread connection is located underneath the unit,
between the two front feet.

Functions

Software function
Press
for at least two seconds to switch the unit on.
The air sampler is supplied with eight programmable air volumes ("Select volume") stored
in the memory and eight Multi Mode configurations selectable by the operator.
The selection of the volume is made by pressing the "UP" or "DOWN" arrows when the
programme in the relevant sub menu.
The following volumes of air are suggested:
Contaminated areas (communities, processing rooms, etc) 10 - 200 litres of air
Normal areas (laboratory benches, houses, etc) 200 - 500 litres of air
Sterile or high risk areas (cleanrooms, operating theatres, etc) 500 - 1000 litres of air
Press
to switch off (from the Main Screen), or each time you need to end an action.