Introduction; Intended Use; Delivery Content And Optional Equipment; Unpacking - VWR Micro Star 30R Instruction Manual

Laboratory centrifuge, inclusive angle rotor ra 24/2 and aerosol tight rotor lid/without rotor
Table of Contents

Advertisement

INSTRUCTION MANUAL

2. INTRODUCTION

2.1 INTENDED USE

Centrifuge Micro Star 30R is a universal laboratory refrigerating centrifuge. It is designed for use
in medical, scientific and industrial laboratories, for separating the substances with different
specific densities, by centrifugal force. In particular, it is intended for preparation of human
samples (body fluids), in approved test tubes, before further analysis. Not intended for in vitro
diagnostic use. The maximum rotational speed of 18,000 RPM gives the centrifugal force of
31,150 xg.

2.2 DELIVERY CONTENT AND OPTIONAL EQUIPMENT

The following is enclosed with the centrifuge:
1 Instruction manual
1 Hexagonal rotor key (magnetic key holder on the back of the device, see page 7)
1 Power cord
1 Angle rotor RA 24/2 with aerosol tight lid for 24 x 1.5/2 ml tubes (only with ECN 521-2574)
Optional rotors and accessories:
521-2576 Angle rotor RA 24/2, 24 x 1.5/2 ml micro tubes with aerosol tight lid, max. RCF
31,150 xg, max. speed 18,000 RPM
521-2577 Angle rotor RA 16/5, 16 x 5 ml micro tubes with aerosol tight lid, max. RCF
21,000 xg, max. speed 14,000 RPM
521-2578 Angle rotor RA 8/15, 8 x 15 ml conical, max. RCF 4,400 xg, max. speed 6,000 RPM
521-2580 Adapter 10 ml, 16 x 100 mm (PK 1) for rotor RA 8/15, max. RCF 4,234 xg
521-2581 Adapter 7 ml, 13 x 100 mm (PK 1) for rotor RA 8/15, max. RCF 4,213 xg
521-2582 Adapter 5 ml, 13 x 75 mm (PK 1) for rotor RA 8/15, max. RCF 3,566 xg
521-2583 Angle rotor RA 6/50, 6 x 50 ml conical, max. RCF 4,400 xg, max. speed 6,000 RPM
521-2585 Adapter 15 ml conical (set of 6), for rotor RA 6/50
521-2586 Swing-out rotor for 4 microplates with 2 carriers, max. RCF 2,200 xg, max. speed
5,000 RPM
521-2587 Swing-out rotor RS 6/12 with buckets, 6 x 12 ml, max. RCF 2,500 xg, max. speed
4,400 RPM

2.3 UNPACKING

The weight of the centrifuge is 50 kg. To prevent possible injuries, at least two people
should lift and carry the centrifuge by holding it at the bottom from opposite sides.
Open the carton box. Take out the accessories and remove the packaging material. Reach with
your hands under the centrifuge and lift it from the box together with another person.
When lifting the centrifuge, never hold it by the front or top plastic part of the
housing or by the lid, as the appliance may get damaged!
Retain the packaging material for any subsequent transport or storage, which are allowed only in
the original packaging.
EN 5 / 27