PerkinElmer TGA 8000 Installation And Hardware Manual page 47

Thermal analysis
Table of Contents

Advertisement

In the Y Data section select the range you plan to use and click OK.
2.
3.
Select Calibrate Instrument from the Calibrate menu.
Select the Weight tab in the Calibration menu.
4.
NOTE: Prior to calibrating the weight the display should look as below with the value in the
Measured(mg) box reading 100.000. If it indicates another value you must restore the
default weight calibration using the Calibration Restore Weight Calibration menu item
Type a name in the Operator ID field (optional).
5.
Enter the weight of the calibration weight in the Ref. Weight field.
6.
NOTE: Remove any pan from the hangdown wire because once you click Begin Calibration you
cannot remove it.
Calibrate the TGA 8000 . 47