GLA531 Series LGR-ICOS™ Gas Analyzer
5 Communications – Data and Alarms

Input/Output Interface Hardware

The input/output (I/O) connections to the analyzer are accessible on the internal interface plate, as shown in
Figure 30 below. Depending on the options installed, the analyzer may have one or all of the connections listed
below:
·
Ethernet (RJ-45)
·
Modbus/TCP (RJ-45)
·
USB
·
4–20 mA analog out (5.08 mm green screw-terminal blocks).
Figure 30 Analyzer I/O Interface Plate
The maximum length
runs are required, use active hardware, such as repeaters or switches, as necessary. See
installation guidelines.
·
For
100BaseTX networking (100 MHz bandwidth), Cat 5 cables with RJ-45 connections are sufficient.
·
For 10GBASE-T (10-Gigabit Ethernet), Cat 6 cables are required for improved performance to 250 MHz
bandwidth.
Table 4 RJ-45 Cable Installation Guidelines from IEEE 802.3u
Parameter
Cable specification
Maximum segment length
Maximum network length
EIA/TIA-568 TSB or EIA-TIA-568 compliant
a
Category 6 SF(Shielded Foil) cable, or 150 ohm STP MII cable in metallic conduit, may be used as alternates
b
3KXG165002R4201_2021_04
of the Ethernet and Modbus cables is limited to
Category 6
a
100 m (328 ft) for 100BaseTX
200 m (656 ft) with one repeater
ABB PROPRIETARY INFORMATION
100 meters per TIA/EIA 568-5-A. If longer
RJ-45 Cable
S/FTP or SF
, 22 to 24 AWG
b
Operating Manual
Table 4
below for cable