Navigator 500
Hydrazine analyzer
7.1.3 Device Setup
Menu
Device Setup
Select
Exit
Menu
Initial Setup
Instrument Tag
Temperature Units
Reset to Defaults
Security Setup
Calibrate Password
Advanced Password
Reset Passwords
Range Change Setup
Range Change
Range 1 Span
Range 2 Span
Range 3 Span
42
This level is used to access standard setup parameters.
Comment
The device (transmitter) identification tag.
Selects the temperature units displayed (°C or °F).
Restores Device Setup parameters to their default settings.
Sets the password to enable access at Calibrate level.
(Not factory-set.)
Sets the password to enable access at Advanced level.
(Not factory-set.)
Clears all passwords.
Notes.
Displayed only if 1 single-stream wet-section is connected.
Range 3 must always be the largest range, Range 1 must always
be the lowest.
Range Changing is disabled automatically when the sensor type
is changed.
When Enabled, Analogue Output 2, Relays 3 and 4 are configured
automatically for range changing and are not available via the
Input/Output sections.
For correct operation, Range 1 Span should be the lowest range and
Range 3 Span should be the highest range.
When Enabled – the following zero and span ranges 1 to 3 are set
automatically:
Range
Zero
1
0
2
0
3
0
The range spans can be changed to suit end-user's requirements.
Span
10
100
1000
Default
Navigator
°C
Disabled
Continued on next page...
OI/AHM550–EN Rev. A