Operating Standards; Unit Specifications - Thermo Scientific TSX Series Installation And Operation Manual

-20 c laboratory and enzyme freezers
Table of Contents

Advertisement

Operating Standards

6 Operating
Standards

6.1 Unit Specifications

Table 2. Unit Specifications
Rated
Model
Voltage
23A / 23Y
115 V
23D / 23Z
208-230 V 2.5 A
23V
208-230 V 2.5 A
30A / 30Y
115 V
30D / 30Z
208-230 V 3.0 A
30V
208-230 V 3.0 A
8
TSX -20°C Laboratory and Enzyme Freezers
The freezers described in this manual are classified for use as stationary
equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental
conditions:
Indoor use
Altitude up to 2000 m (6512 feet)
Maximum relative humidity 60% for temperatures from 15 to 32°C
(59 to 90°F).
Main supply voltage fluctuations not to exceed 10% of the nominal
voltage.
The freezer must not be connected to a GFCI (Ground Fault Circuit
Interrupter) protected outlet as it may be subject to nuisance tripping.
Operation outside of these conditions will affect the performance of the unit
and the samples stored inside.
The last character in the model number listed on the freezer identifies the
electrical specifications for your unit.
The dataplate is located on the upper left side of the unit.
Rated
Frequency
Current
/ Phase
4.3 A
60 Hz/1
50/60 Hz/1
50/60 Hz/1
4.70 A
60 Hz/1
60 Hz/1
50 Hz/1
Power
Approximate
Module
Lab Freezer
Plug
Weight Kg (lbs)
178 (394)
169 (373)
IEC C19
188 (415)
Approximate
Exterior
Enzyme Freezer
Dimensions
Weight Kg (lbs)
(D x W x H)
203 (449)
93.6 x 71.1 x 199.4 cm
194 (428)
(36.9 x 28.0 x 78.5 in)
99.0 x 86.4 x 199.4 cm
220 (484)
(38.9 x 34.0 x 78.5 in)