Thermo Scientific TSX Series Installation And Operation Manual page 14

-20 c laboratory and enzyme freezers
Table of Contents

Advertisement

Installation
10
TSX -20°C Laboratory and Enzyme Freezers
Table 3. TSX -20°C Freezer Model Numbers Requiring Anti-Tip
Bracket Installation*
TSX2320FV
TSX2320EV
TSX2320FY
TSX2320EY
TSX2320FZ
TSX2320EZ
*Check the product data plate to confirm model number.
Anti-tip Bracket Installation instructions are provided for wood and concrete
floors. Any other type of construction may require special installation techniques
as deemed necessary to provide adequate fastening of the anti-tip bracket to the
floor. For installation on floors other than wood and concrete, please contact
technical support.
The use of this bracket does not prevent the tipping of the freezer when not
properly installed.
Figure 1. Materials Supplied
Label
Description
1
Bracket
2
Bolts
3
Anchors
4
Instructions and Installation Template
Table 4. Tools required
Wood Floor
Flashlight
Tape Measure
1/2
(13 mm) Wrench
"
3/4
"
(19 mm) Wrench
Drill
15/64
(6 mm) Drill Bit
"
TSX3020FV
TSX3020FY
TSX3020FZ
Concrete Floor
Hammer Drill
1/2
(13 mm) Masonry Bit
"
TSX3020EV
TSX3020EY
TSX3020EZ
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents