Operating Standards; Electrical Specifications - Thermo Scientific TSX Series Installation And Operation Manual

Ultra low temperature freezers
Table of Contents

Advertisement

6 Operating
Standards
6.1 Electrical
Specifications
Thermo Scientific TSX Ultra Low Temperature Freezers
The freezers described in this manual are classified for use as stationary
equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental
conditions:
Indoor use
Altitude up to 2000m
Maximum relative humidity 60% for temperatures up to 32°C (90°F)
Main supply voltage fluctuations not to exceed 10% of the nominal
voltage.
The ULT should not be connected to a GFCI (Ground Fault Circuit
Interrupter) protected outlet as it may be subject to nuisance tripping.
The last character in the model number listed on the dataplate identifies the
electrical specifications for your unit.
The voltage type is, as specified in the following table:
Table 1. Electrical Specifications
Model
Voltage
V
230V
Frequency
50Hz
Installation and Operation Manual

Operating Standards

Current
4A
7