Operating Standards

The freezers described in this manual are classified for use as
stationary equipment in a Pollution Degree II and over voltage
Category II environment.
These units are designed to operate under the following
environmental conditions:
Indoor use.
Altitude up to 2000 m.
Maximum relative humidity 60% for temperatures within
15 °C to 32 °C (59 °F to 90 °F).
Main supply voltage fluctuation shall not exceed ±10% of
the nominal voltage. For the 100V input condition this
should be +10% and -8%.
For the TSX series, the ULT should not be connected to a
GFCI (Ground Fault Circuit Interrupter) protected outlet as
it may be subject to nuisance tripping.

Electrical Specifications

The data plate identifies the electrical specifications and
current ratings for your unit.
Table 1. TSX Series Electrical Specifications
Model
Voltage
TSX40086FA
TSX40086LA
TSX50086FA
TSX50086LA
100-230 V
TSX60086FA
TSX60086LA
TSX70086FA
TSX70086LA
Note: Specified voltage range implies the input voltage is a
standard global voltage.
8 |
Operating Standards
Frequency
Current
50/60 Hz
8.9-4.9 A
Ultra Low Temperature Freezers