7.2 Wiring

Thermo Scientific
The wiring diagrams are attached on the back of the cabinet.
CAUTION: Connect the equipment to the correct power source. Incorrect
voltage can result in severe damage to the equipment.
CAUTION: For personal safety and trouble-free operation, this unit must
be properly grounded while in use. Failure to ground the equipment may
cause personal injury or damage to the equipment. Always conform to the
National Electrical Code and local codes. Do not connect the unit to
overloaded power lines.
CAUTION: Do not position the unit in a way that impedes access to the
disconnecting device or circuit breaker in the back of the unit.
CAUTION: Always connect the unit to a dedicated (separate) circuit. Each
unit is equipped with a service cord and plug designed to connect it to a
power outlet which delivers the correct voltage. Supply voltage must be
within
10% of the unit rated voltage. If cord becomes damaged, replace
±
with a properly rated power supply cord.
Table 5. Power Cord Specifications
Model
Power Cord Specifications
A / Y
3-G 12 AWG, NEMA 5-15P, 15 A/125 V
D / Z
3-G 14 AWG, NEMA 6-15P, 15 A/250 V
V
3-G 1.5 mm
CAUTION: Never cut the grounding prong from the service cord plug. If
the prong is removed, the warranty is invalidated.
CAUTION: In an emergency, the power cord is a disconnect device.
The Laboratory and Enzyme freezer has a DeviceLink HUB connector. Refer to
instructions provided with the DeviceLink HUB guide.
2
, CEE 7/7, 16 A/250 V
TSX -20°C Laboratory and Enzyme Freezers