Troubleshooting
Table 11. Troubleshooting Procedures
Problem
Cause
Unit is warm
around door
Perimeter Heater is ON
frame
Door is open
Door seal
Warm product recently
loaded in unit
Unit warms up
Power supply
Setpoints need to be
adjusted
"E01" on display
Invalid Algorithm
"E02" on display
Control Probe Failure
"E05" on display
Ambient Probe Failure
Compressor Run Time
"E07" on display
High
Upper Bottle Probe
"Err" on display
Failure
"---" on display
Lost Communication
40
TSX -20°C Laboratory and Enzyme Freezers
Solution
This is a normal function of the unit and is a result of the perimeter heater to
reduce condensation.
Make sure the door is completely closed.
Check the door seal, following instructions in
Allow ample time to recover from loading warm product.
Check for proper voltage to the unit. If there is no voltage to the unit, call an
electrician.
To adjust the setpoint, refer to
Check to ensure the model type is set correctly in service mode. Refer to
.
Section 9.4
Check for loose probe connector. Replace control probe.
Check for loose probe connector. Replace ambient probe.
The compressor is working harder to overcome the ice on the unit walls and
may start to thermally fail. Defrost the unit.
Check for loose probe connector. Replace upper bottle probe.
Call customer service.
Section 12.5
Section 9.3
.
.