Thermo Scientific TSX Series Installation And Operation Manual page 15

-20 c laboratory and enzyme freezers
Table of Contents

Advertisement

Thermo Scientific
1. Locating the Bracket
a. Determine where you want the centerline of the freezer to be.
b. Place the included template on the floor lined up with the centerline
of the freezer and keep 6"- 12" between the wall and the back of the
unit.
c. On the floor, mark the location of Hole #1 & Hole #2.
2. Anti-Tip Bracket Installation
Wood Construction
a. Drill 15/64" (6 mm) pilot holes in locations marked in step 1.
b. Place bracket on floor aligned with holes.
c. Use supplied lag bolts to attach bracket to floor.
Concrete Construction
a. Drill 1/2" (13 mm) holes in locations marked in step 1 with
masonry bit.
b. Slide lag screw anchors into holes to be flush with floor surface.
c. Place bracket on floor aligned with holes.
d. Use supplied lag bolts to attach bracket to floor.
3. Adjusting Bolt in freezer
a. Locate 1/2" bolt attached to bottom of cabinet.
b. Unscrew 1/2" bolt until there is 1/2" clearance between floor and
head of bolt as shown in Figure 2.
c. Tighten lock nut against bottom of unit.
TSX -20°C Laboratory and Enzyme Freezers
Installation
11

Hide quick links: