7 6 water Supply Line Connection

• Grasp the Water Supply Line (blue hose) by the end
opposite the quick-disconnect and insert it into the water
inlet connector until fully seated.
• Connect the quick-disconnect to the dental office water
supply or a Cavitron DualSelect Dispensing System.
• Inspect all connections to make certain there are no
leaks.
• To remove the water line from the Cavitron System, turn
off the dental office water supply. Disconnect the water
supply line from the dental office water supply. If a quick-
disconnect connector is attached to the end of the hose,
relieve the water pressure by pressing the tip of the
connector into an appropriate container and allow water
to drain. To remove the hose from the system, push on
the outer ring of the system's water inlet and gently pull
out the water line.

7 7 TAp-ON

FOOT pEdAL

Tm

SYNChrONizATiON

The Tap-On Wireless Rechargeable Foot Pedal supplied
with your system has been factory synchronized with the
base unit. Should a replacement Foot Pedal be necessary,
synchronization will be required prior to system operation.
Perform the following steps to synchronize the Foot Pedal
with the base unit.
1. Turn the Main Power switch located at the rear of the
system to the ON (I) position. The Main screen will appear.
This is also referred to as the Scale screen.
2. Press and hold the Settings icon until the Settings screen
appears.
3. Maintain a distance of no more than 10 feet (3 meters)
between the base unit and Tap-On Foot Pedal during the
synchronization process.
4. Remove any inserts from the handpiece.
5. Ensure the foot pedal power is ON.
6. Press the Sync icon on the Settings screen.
7. Press the red sync button on the bottom of the foot pedal -
hold for at least three seconds.
8. Synchronizing is complete when the Sync icon has
9. To verify proper communication, press the Scale icon to
7 8 rEChArgiNg ThE TAp-ON FOOT
pEdAL
Press ring to
Ensure the foot pedal power is ON. With the unit ON, plug
release water
supply tube.
the Auxiliary foot pedal cable into the foot pedal and the USB
port on the front of the unit. Allow to charge for 2 hours for a
full battery charge.
The unit will operate with the foot pedal while the battery is
charging.
Battery Icon (Foot Pedal Charge)
7 9 pOwEriNg ON/OFF ThE TAp-ON
FOOT pEdAL
• The foot pedal is packaged with the power OFF. The foot
• The Power button is used to disconnect the battery from
• The Power button must be pressed and held for 4
• If the Foot Pedal power is OFF, pressing the Sync button
6
stopped rotating and sound indication occurs.
return to the Main screen (scale screen). Depress the foot
pedal to the second position and ensure that Boost is
activated (seen in Power Level icon).
pedal must be turned on prior to use. (The foot pedal will
not charge in the OFF state).
the circuitry for shipping purposes.
seconds to disconnect the battery. Likewise the Power
button must be pressed and held for 4 seconds to
reconnect the battery.
or trying to Sync with the Scaler will not turn the Foot
Pedal on.