• As a medical device, this product must be installed
in accordance with applicable local, regional, and
national regulations, including guidelines for water quality
(e.g. drinking water). As an open water system, such
regulation may require this device to be connected to
a centralized water control device. The Cavitron
DualSelect Dispensing System may be installed to allow
this unit to operate as a closed water system.
SECTiON 7:
installation instructions
Anyone installing a Cavitron Touch System should observe
the following requirements and recommendations.
7 1 water Line requirements
• A water supply line with user-replaceable filter is
supplied with your system. See Section 10 System Care
for replacement instructions.
• Incoming water supply line pressure to the system must
be 20 psi (138 kPa) to 40 psi (275 kPa). If your dental
water system’s supply line pressure is above 40 psi,
install a water pressure regulator on the water supply line
to your Cavitron Touch Ultrasonic Scaler.
• A manual shut-off valve on the dental water system
supply line should be used so that the water can be
completely shut-off when the office is unoccupied.
• In addition to the water filter supplied, it is recommended
that a filter in the dental water system supply line be
installed so that any particulates in the water supply will
be trapped before reaching the Cavitron system.
• After the above installations are completed on the dental
water supply system, the dental office water line should
be thoroughly flushed prior to connection to the Cavitron
system.
• Incoming water temperature to the Cavitron System
should not exceed 25˚C (77˚F). If needed a device
should be installed to maintain a temperature within this
specification, or a Cavitron DualSelect Dispensing
System attached to allow this system to be operated
as a closed water system.
7 2 Electrical requirements
• Incoming power to the system must be 100 volts AC to
240 volts AC, single phase 50/60 Hz capable of
supplying 1.0 amps.
• The system power should be supplied through
the AC power cord provided with your
system.
• WARNING: To avoid risk of electric shock, this equipment
must only be connected to a supply mains with
protective earth.