Operation; Accessories And User; Replaceable Parts; Accessories - DENTSPLY Cavitron Touch Directions For Use Manual

Ultrasonic scaling system
Hide thumbs Also See for Cavitron Touch:
Table of Contents

Advertisement

8 5 Tap-On Technology wireless Foot
pedal Operation
Using Foot Pedal in Tap-On Mode
For scaling procedures,Tap-On Technology eliminates the
need to hold the pedal down. Tapping the foot pedal once
activates ultrasonic power for approximately 4 minutes.
Tapping the foot pedal while in Tap-On mode disables the
ultrasonic power and water flow. Boost is still available while
scaling in Tap-On mode. To use Boost, simply depress the
foot pedal to the second position (all the way to the floor) to
activate and hold as long as Boost is desired. Release foot
pedal to return to Tap-On mode.
DEPRESSED
NON-DEPRESSED
1st POSITION
TIPS:
Tap-On Technology will not run water unless an
insert is in the handpiece
A sensor in the handpiece holder will prevent
Tap-On Technology from operating when the
handpiece is in the holder.
If the foot pedal is not tapped quickly, it will
function in a conventional manner.
How to Enable and Disable Tap-On Technology
The Tap-On Technology feature can be enabled by pressing
the Tap-On icon in the Settings Screen. From the main
screen, press the Settings icon then press the Tap-On icon.
The parenthesis lines around the icon with show as white
when Tap-On is activated and when you return to Scale (Main
screen) TAP ON will appear in the top of the Power Level
bubble.
The Tap-On Technology feature can be disabled by pressing
the Tap-On icon in the settings. From Scale (Main Screen)
press the Settings icon then press the Tap-On icon. The icon
will show as grayed out.
Using Foot Pedal without Tap-On Mode
For scaling operation, the first position activates both the
ultrasonic energy and lavage at the insert tip. The second
position activates the Boost Mode. The Boost Mode (fully
depressed Foot Pedal) increases the ultrasonic power level
for quick removal of tenacious deposits without adjusting the
power level knob. To deactivate Boost Mode, release the
Foot Pedal to the first position.

8 6 Accessories and user

replaceable parts

8 6 1 Accessories

1. AC Power Cord
2. Wireless, Rechargable Foot Pedal (with Tap-On Technology)
3. Auxiliary Foot Pedal Power Cable
®
4. Cavitron Steri- Mate
360 Sterilizable Handpiece
5. Cavitron 30K Ultrasonic Inserts
6. Cavitron DualSelect Dispensing system
7. Detachable Featherweight Handpiece Cord
8. Cavitron Steri-Mate Sterilizable Handpiece
8 6 2 user replaceable part Kits
1.
2.
3.
For detailed information, contact your local DENTSPLY
Representative or authorized DENTSPLY Distributor.
SECTiON 9: System Setup,
DEPRESSED
Operation and Techniques
2nd POSITION
for use

9 1 handpiece Setup

• Hold empty handpiece in a semi-upright position over a
• Lubricate the O-ring on the insert with water before
• Turn the Lavage Control to select flow rate during
11
Cavitron Insert Replacement O-ring Kits, 12/packs
Part Number 62351 (black) for plastic and soft grips
Part Number 62605 (green) for metal grips
Handpiece Cable O-Ring, Part Number 79357
Lavage (Water) Filter, 10/Pack, Part Number 90158
This handpiece is sterilizable. Refer to Cavitron Systems
Infection Control Procedure booklet for sterilization
instructions prior to using handpiece.
Connect the Handpiece to the Cable Assembly by
aligning the electrical connections. If Cable Assembly
does not seat into the handpiece, gently rotate
the handpiece until contacts align, then fully insert
handpiece. NOTE: Steri-Mate 360 and Steri-Mate are
compatible with this system.
sink or drain. Activate the Tap-On Foot Pedal until water
exits to release any air bubbles that might be trapped
inside the handpiece. NOTE: Tap-On Technology only
operates when an insert is in the handpiece.
placing it into the handpiece. Fully seat insert with a
gentle push-twist motion. DO NOT FORCE.
system operation. Flow rate is placed on a scale from 1
to 6. Turn control clockwise toward 6 to increase flow
at insert tip. Turn control counter-clockwise toward 1 to
decrease flow. The flow rate through the handpiece also
determines the temperature of the lavage. Lower water
flow rates produce warmer lavage. Higher flow rates
produce cooler lavage. If the handpiece becomes warm,
increase the flow rate. With experience the Dental
Healthcare Professional will be able to determine the
best flow rate setting for optimum operating efficiency
and patient comfort.