8 3 handpiece / Cable
Lavage Control
Turn the Lavage Control to select flow rate during system
operation. Flow rate is based on a scale from 1 to 6. Turn
clockwise toward 6 to increase flow at insert tip. Turn
counter-clockwise toward 1 to decrease flow. The flow rate
through the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer lavage. Higher
water flow rates produce cooler lavage.
360° Rotating Handpiece Nose
To rotate the insert, place fingers on the nose of the
handpiece and rotate to desired position. Handpiece
maneuverability allows adjustable hand positioning, free
flowing movement and access within the anterior and
posterior of the oral cavity.
Note: The design of the Steri-Mate 360 provides a smooth
transition to Cavitron FitGrip inserts (shown).
8 4 Cavitron 30K ultrasonic inserts
The many styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications.
O-Ring
Provides seal for handpiece coolant.
O-ring should be replaced when worn.
Insert Tip
Shape and size of tip
determine access and
adaptation. Preheated
Lavage directed to tip.
Insert Marking
Finger Grip
Manufacturer,
Date (Julian date code
YDDD), Frequency,
Type, Tip Lot Number
(if applicable).
Featherweight, Detachable Handpiece Cord
Lightweight handpiece cable easily detaches from the
unit, also detaches from the Steri-Mate 360 handpiece
for cleaning and replacement needs. Swivel feature
reduces cord drag as handpiece rotates during
procedures.
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the handpiece into
mechanical oscillations used to activate the insert
tip.