30 Using the RetinaVue 100 Imager Pro
4. Touch Start to enter the Image acquisition mode and begin the first exam of the
patient's right eye (R1).
The Exam acquisition screen appears.
5. Hold the patient end of the RetinaVue 100 Imager Pro two to three inches directly in
front of the patient's examined eye. Continue forward to compress the eye cup
against the examined eye.
6. Direct the patient to focus on the green fixation lights inside the barrel of the
RetinaVue 100 Imager Pro.
Note
WARNING Clean and disinfect the eye cup after each
patient to avoid the risk of cross-contamination.
Instruct the patient to cover, but not close, their unexamined
eye. This will help the patient to focus on the green fixation
lights.
Welch Allyn RetinaVue™ 100 Imager Pro