Digistar SCALE LINK SL2 Technical Manual page 38

Table of Contents

Advertisement

Output example: Scales A & B with scale B selected at the indicator.
1
12345678901234567890123456789012345678901234567890
"^
The weight tag changes from (NE,GR, or LU) to (NC,GC,or LC) when indicator is performing internal temperature
calibrations (which can take up to 3 seconds). The weight sent at this time is the same weight prior to internal
temperature calibrations. This occurs every 20 minutes but is done several times during the first 20 minutes after
power up.
Output example: Scales A & B with scale B selected at the indicator. Both scales are being calibrated.
1
12345678901234567890123456789012345678901234567890
"^
Weight will appear as "999999ES" when operator is in Short/Long Form Menu's.
Output example: Scales A & B with scale B selected at indicator, B scale is in setup.
1
12345678901234567890123456789012345678901234567890
"^
Weight will appear as "999999ER" when weighing errors (+/-RANGE, CHK AD, OVERCAP etc..) are
detected by the scale.
Output example: Scales A & B with scale B selected at the indicator, B scale has an error.
12345678901234567890123456789012345678901234567890
"^
280LB GR,> 999999LB ER@C~"
38
Transmit comma delimited data with basic weighing information for each scale platform one time per
second.
Same data format as setting 37.
39 Transmit comma delimited data with basic weighing information for each scale platform ten times per
second.
Same data format as setting 37.
38
2
3
280LB GR,>
11300LB NE@C~"
2
3
280LB GC,>
11300LB NC@C~"
2
3
280LB GR,> 999999LB ES@C~"
1
2
Scale Link 2 Technical Manual
4
5
4
5
4
5
3
4
5
D4204-EN