Getting Started; Initial Set-Up; Language Selection - BD Alaris 8002TIG03 Directions For Use Manual

Syringe pump with plus software mk4
Hide thumbs Also See for Alaris 8002TIG03:
Table of Contents

Advertisement

Getting Started

Initial Set-up

Before operating the Pump read this Directions For Use manual carefully.
w
1 . Check that the Pump is complete, undamaged and that the voltage rating specified on the label is compatible with your AC power
supply .
2 . Items supplied are:
• Alaris Syringe Pump
• User Support CD (Directions For Use)
• AC Power Cable (as requested)
• Protective Packaging
3 . Connect the Pump to the AC power supply for at least 2½ hours to ensure that the internal battery is charged (verify that the S is
lit) .

Language Selection

1 . On initial start-up the Pump will display the Select Language screen .
2 . Select the required language from the list displayed using the
3 . Press the OK softkey to confirm your selection .
• The Pump may be used safely with the pre-installed default data set. Any data set created for installation must be
w
approved by an appropriately qualified person with clinical authority in accordance with hospital protocol prior
to upload and activation.
• The Pump will automatically operate from its internal battery if the Pump is switched on without being connected
to the AC power supply.
• Should the Pump fail to perform correctly, replace in its original protective packaging, where possible and contact
Qualified Service Personnel for investigation.