Getting Started; Initial Set Up; Power Input - BD Alaris GW 800 Series Directions For Use Manual

Volumetric pumps
Hide thumbs Also See for Alaris GW 800 Series:
Table of Contents

Advertisement

Getting Started

Before operating the Pump read this Directions For Use (DFU) manual carefully.
w

Initial Set Up

1 . Check that the Pump is complete, undamaged and that the voltage rating specified on the label is compatible with your AC power
supply .
2 . Items supplied are :
• Alaris GW 800 Volumetric Pump
• User Support CD (Directions For Use)
• AC Power Cable (as requested)
• Protective Packaging
• Model 180 Flow Sensor (Pump model dependent, if not supplied then the flow sensor can be ordered separately)
• Electronic Instructions For Use Insert
3 . Connect the Pump to the AC power supply for at least 2½ hours to ensure that the internal battery is charged (verify that the
is lit) .
The Pump will automatically operate from its internal battery if the Pump is switched on without being connected
w
to the power supply.
Should the Pump fail to perform correctly, replace in its original protective packaging, where possible and contact
Qualified Service Personnel for investigation.
4 . Facilities intending to use non-dedicated infusion sets with the Set Detection Mode disabled are advised to assess performance
of the Pump and infusion set combination prior to clinical use . If adequate performance is not achieved, then the Alaris GW 800
branded dedicated infusion sets should be used or the Pump should be recalibrated by following the volumetric calibration
procedure described in the Technical Service Manual Addendum 1000SM00014 .
Factory set calibration values in the Alaris GW 800 Volumetric Pump are intended specifically to support the use
w
of Alaris GW 800 branded dedicated infusion sets. Compatibility of factory calibration settings with Alaris GW 800
branded dedicated infusion sets is controlled by BD.
BD reserves the right to make changes to the Alaris GW 800 Volumetric Pump without notice which may affect
its performance with non-dedicated infusion sets. If adequate performance is not achieved, then Alaris GW 800
branded dedicated infusion sets should be used or the Alaris GW 800 Volumetric Pump should be re-calibrated.
It is recommended that the performance of any non-dedicated infusion set is re-evaluated periodically as the set
manufacturer may change specifications significant to the performance of the Alaris GW 800 Volumetric Pump
without notice.

Power Input

The Pump is powered from the AC supply through a standard IEC AC connector . When connected to the AC supply the AC Power
indicator is illuminated .
• To isolate the Pump from AC supply remove the AC connector from the source socket.
w
• The Pump should be positioned to allow access for disconnecting the AC connector.
BDDF00160 Issue 1
Alaris™ GW Volumetric Pumps
11/40