Add Drug

This option allows the user to add a drug whilst the Pump is running a ml/h infusion .
Note: This option will not be available when operating the Alaris CC Syringe Pump or the Alaris CC Guardrails Syringe Pump in semi
dedicated mode and using an extension set without pressure disc .
d
1 . Press the
button to access the options menu .
2 . Select ADD DRUG option using the
3 . Select from Drug from the displayed list using the
NOTE: Drugs are configured via Alaris Plus Editor software based on the units available (time / weight based):
w
• Gram base: ng, mcg/μg, mg and g
• Unit base: mU/mUnits, U/Units and kU/kUnits
• mmol base: mmol
• Volume base: ml (Warning: If a running ml/h infusion is promoted to a drug/fluid with dosing unit of ml, from the
drug library using the Add Drug functionality, the newly promoted drug/fluid will not contain any predefined soft
or hard limits).
4 . Select drug name using the
5 . Select concentration using the
6 . Select patient weight using the
7 . Confirm drug, overriding Guardrails soft limits*, as appropriate .
Guardrails
*

Adjust Alarm Volume

To change the alarm volume, if enabled .
d
1 .
Press the
button to access the options menu .
2 . Select ADJUST ALARM VOLUME .
Note: The Pump will beep at the selected alarm volume setting . The user must assess whether the alarm volume setting is loud
enough for the intended operating environment, and adjust appropriately .
3 . Select alarm volume required and press the OK softkey .

Profile Filter

Guardrails
Configure the profiles to be enabled or disabled in the list of selectable profiles, if enabled .
1 . Stop the infusion . Press the
2 . Select PROFILE FILTER .
3 . Select Profile(s) required to change and press the MODIFY softkey .
4 . Press the OK softkey to confirm .

Standby

This option allows the Pump to be placed on standby mode, if enabled .
1 . Stop the infusion . Press the
2 . Select STANDBY using the
3 . Select CANCEL to return to main display .
f
keys and press the OK softkey .
f
f
keys, press OK to confirm .
f
keys, press OK to confirm, if applicable .
f
keys, press OK to confirm, if applicable .
d
button to access the options menu .
d
button to access the options menu .
f
keys, press the OK softkey to confirm .
BDDF00001 Issue 5
keys, press OK to confirm .