BD Alaris 8002TIG03 Directions For Use Manual page 31

Syringe pump with plus software mk4
Hide thumbs Also See for Alaris 8002TIG03:
Table of Contents

Advertisement

Display
ADD DRUG NOT COMPLETE
ATTENTION
*Configurable option .
Note: The audio sound pressure level is at least 45 dB depending on configuration of the alarm sound level .
Setting the alarm sound pressure level lower than the ambient sound pressure level can impede User recognition of
w
alarm conditions.
Alarm Priority Level Indicators
Priority
Audio Indicator
HIGH
Ten beep sequence followed by a three second pause
MEDIUM
Three consecutive beeps followed by a four second pause
LOW
Three consecutive beeps followed by a sixteen second pause
Alarm
Description and Troubleshooting Guide
Priority
The ADD DRUG operation has not been completed after five seconds the User will be
Low
notified by an auditory cue . After two minutes have expired and the ADD DRUG operation
has not been completed, a low priority alarm is generated . Press CANCEL softkey and
complete ADD DRUG operation .
Low
If the Pump has been left on for more than 2 minutes* (referred to as ATTENTION in the
log) without starting the operation, a low priority alarm will be generated . Press the
button to silence the alarm for a further 2 minutes . For extended Attention Timeout press
and hold down the
Pump on standby for 15 minutes .
BDDF00001 Issue 5
R
button and wait for four beeps in succession, this will put the
30/56
Alaris™ Syringe Pumps
Alarms and Warnings
R
Visual Indicator (Beacon)
Flashing Red
Flashing Amber
Solid Amber