Alaris Editor Software Profile Configuration - BD Alaris 8002TIG03 Directions For Use Manual

Syringe pump with plus software mk4
Hide thumbs Also See for Alaris 8002TIG03:
Table of Contents

Advertisement

General Options
1 . Select GENERAL OPTIONS from the Configured Options menu using the
2 . Select the option required to enable/disable or adjust and press the MODIFY softkey .
3 . When all the required modifications have been carried out press the QUIT softkey .
4 . Either select the next configuration option from the menu or turn the Pump off, returning it to operation as required .
NURSE CALL FITTED
NURSE CALL INVERT
RS232 SELECTED
DOUBLE DECIMAL ML/H
REMINDER SIGNAL

Alaris Editor Software Profile Configuration

The following options are only configurable via the Alaris Editor Software (PC based), see Alaris Editor Directions For Use for details on
how to configure Profile Configurations .
Data Set Configuration Settings
Hospital Name
Profile Filter
Guardrails
Microgram
Unit Display
Unit
General Pump Configurations
AC Fail
Audio Volume
Audio Volume Adjustable
Auto Night Mode
Auto Save
Battery Icon
Callback Time
Drug Override Mode
Guardrails
Event Log
Pressure Display
Quiet Mode
Rate Titration
Rate Lock
Standby Mode
VTBI Clear Rate
Weight Default
Weight Soft Minimum
Guardrails
Weight Soft Maximum
Guardrails
Enables Nurse Call (hardware option) .
When enabled, the nurse call output is inverted .
Sets the Pump's communications to use RS232 (hardware option) . The NURSE CALL FITTED option must
be enabled to allow RS232 to be enabled .
When enabled, the ml/h rate, VI and VTBI will be displayed to two decimal places . When disabled, the
ml/h rate, VI and VTBI will be displayed to one decimal place .
When enabled there is an audible notification, consisting of four beeps, every 10 minutes for the
Low Battery and Near End Of Infusion alarms .
Note: Reminder Signals will only function when 3RD EDITION ALARMS are enabled .
Configure the name of the facility to be displayed on the Pump .
Controls whether the user is able to filter which profiles will be available on the Pump .
The text used to display microgram, either mcg or µg .
The text used to display Units, either mU, U and kU or mUnit, Unit and kUnit .
The AC Power Failure Alarm can be set to sound or be silent if the AC power is removed .
The audio alarm volume of the Pump (high, medium or low) .
Sets whether the user is able to adjust the audio volume setting .
Main Display (Backlight) dims between hours 21:00 and 06:00 .
Feature to retain previous settings when Pump is switched on .
Indicator displaying the remaining estimated battery capacity .
Adjusts the length of time before the Pump sounds the Attention alarm .
Always - Confirmation of setting will be required for any changes made to the dose rate that are
outside of the Guardrails Soft Alerts
Smart - Confirmation of setting will be required on first dose rate set outside of the Guardrails Soft
Alert . Any subsequent changes will not require confirmation until after the dose rate has been
confirmed inside the Guardrails Soft Alert limits . Additionally any changes in dose rate from above a
Soft Alert Max to below a Soft Alert Min or from below a Soft Alert Min to above a Soft Alert Max will
also need to be confirmed .
The event log can be set to be displayed or not on main display . Events are still recorded in the log .
Sets whether the Pressure Information is available on the display .
Mode to silence key press tones and power down sequence .
Feature to adjust the infusion rate while the Pump is infusing, without putting the Pump on hold .
Anti-tamper feature which prevents rate changes, bolus operations and powering Pump down .
Sets whether the Standby Mode is available on the Pump .
Infusion rate will be set to zero when VTBI has been completed .
The default patient weight in kg .
The minimum patient weight in kg . This is a Guardrails Soft Alert and can be overridden .
The maximum patient weight in kg . This is a Guardrails Soft Alert and can be overridden .
BDDF00001 Issue 5
f
keys and press the OK softkey .
33/56
Alaris™ Syringe Pumps
Configured Options