General Syringe Pump Configurations
Back Off
Display Syringe Brand
Manual Bolus
Bolus Mode
Bolus Rate Default
Bolus Rate Max
Bolus Volume Max
Infusion Rate Max
Near End of Infusion Point
End of Infusion
KVO at EOI
KVO Rate
Purge Syringe
Purge Rate
Purge Volume Max
VTBI Max
Pressure Maximum
Occlusion Alarm Pressure
Auto Pressure
Auto Set Pressure
Auto Offset
GH
Pressure Maximum
GH
Occlusion Alarm Pressure
The approved data set contains configurable option values per profile.
w
An automatic feature which is activated following an occlusion . The Pump action reverses and
pumps backwards to release the pressure which has built up in the infusion system, this minimises
the post occlusion bolus .
Sets whether the syringe brand and size is shown while Pump is infusing .
Bolus delivered by manually moving the plunger mechanism during an infusion or while on hold .
Volume infused displayed will be increased accordingly .
Bolus feature can be set to one of the following options:
• Disabled
• Hands-On only
• Hands-Free and Hands-On
The default value for bolus rates .
The maximum value for bolus rate .
The maximum permissible bolus volume .
The maximum value for infusion rate .
Sets the Near End Of Infusion warning time, as time left to End Of Infusion .
Sets the End Of Infusion point, as a percentage of syringe volume .
Sets whether the Keep Vein Open (KVO) at End Of Infusion (EOI) is available .
Sets the Keep Vein Open (KVO) rate at which the Pump will operate when End of Infusion (EOI) is
reached .
Feature which prompts the user to purge the extension set prior to the start of the infusion .
The rate used during purge operation .
The maximum permissible purge volume .
The maximum value for Volume To Be Infused (VTBI) .
The maximum occlusion pressure alarm value that can be selected during an infusion .
The default occlusion pressure alarm value that can be selected during an infusion .
Feature to set the occlusion pressure alarm level to an amount (mmHg) above the current in-line
pressure, using a single key press .
Automatic feature to set the occlusion pressure alarm level to an amount (mmHg) above the current
in-line pressure, 15 minutes after starting the infusion .
The automatic offset value in mmHg used by auto pressure and auto set pressure .
The maximum occlusion pressure alarm level that can be selected during an infusion .
The default occlusion pressure alarm level that can be selected during an infusion .
BDDF00001 Issue 5